Broadley Technologies 14 Manual de Instruções Página 1

Consulte online ou descarregue Manual de Instruções para Medição, teste e controlo Broadley Technologies 14. Broadley Technologies 14 Instruction manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Transmitter

• Wiring DiagramPage 3-2• Programmingthe DisplayPage 5-13• CalibrationChapter 6• TroubleshootingChapter 7QUICK GUIDEInstruction Manual19 Thomas, Irvin

Página 2 - Table of Contents

Panel mountingPanel cut-out: 138 mm x 138 mm (5.4” x 5.4”)Front panel dimensions: 144 mm x 144 mm (5.7” x 5.7”)Included hardware: 2 flat head screws

Página 3

Model 14 Oxygen AnalyzerVertical pipe mounting Included hardware: 2 pan head screws, 4 mm x 60 mm long for red clamping bow Horizontal pipe mounting

Página 4 - Introduction

Model 14 Oxygen AnalyzerChapter3: Installation of the Instrument3.1 UnpackingThe analyzer should be unpacked with great care.Watch for any loose acces

Página 5

Model 14 Oxygen Analyzer3.8 Electrical Connections Model 14Do not switch the instrument on until completion of the installation.An aluminum armor pla

Página 6

Model 14 Oxygen Analyzer3.9 Terminal Descriptions3-3DESCRIPTION CONNECTION0 - 20 mA or 4 - 20 mA (n°1) [+] user 0 - 20 mA or 4 - 20 mA (n°1) [-] use

Página 7

Model 14 Oxygen Analyzer3-43.10 Changing the Programming LanguageThe default programming language is English. To change the language, follow the proce

Página 8 - Description of the Analyzer

Model 14 Oxygen Analyzer3-53.11 Programming the Power Supply FrequencyThe power supply frequency can be changed if necessary. This change occurs at th

Página 9

Model 14 Oxygen Analyzer3-6

Página 10 - Model 14 Oxygen Analyzer

Model 14 Oxygen AnalyzerChapter4: Using the Instrument4.1 Front Panel KeysThe display may be programmed to indicate : Sample concentration Sample te

Página 11

Model 14 Oxygen Analyzer4-24.2 Displays 1 to 4 (live displays)100% : Dissolved Oxygen measurement 23.2°C : Temperature measurementO2 : ApplicationS1..

Página 12 - Chapter3:

Model 14 Oxygen AnalyzerTable of ContentsiChapter 1 : Introduction ...

Página 13

Model 14 Oxygen Analyzer4.3 Description of the Function KeysThe function keys below will be highlighted at the bottom of the screen :4-34.4 IconsWait

Página 14

Model 14 Oxygen Analyzer4.5 Enter or Modify a ValueThe highlighted digit can be modified with the key .Each digit can be validated by pressin

Página 15 - 23.2°CO2

Model 14 Oxygen AnalyzerChapter5: Programming the Transmitter5.1 Main Menu5-110023.2°CO2% SatDisp2 MenuS1 S3S2 S4CONC. CALIB.TEMP. CALIB.PARAMETERSCAL

Página 16

Model 14 Oxygen Analyzer5.1.1 CALIBRATION Menus➯ Any calibration should follow the procedure below:1. Configure the calibration characteristics in the

Página 17

Model 14 Oxygen Analyzer5.1.2 MAINTENANCE Menu5-321.6°C60.5 nA21.0 ppmMAINTENANCEAn access code may be required.See page 5-14 for CODE menu.mUsed for

Página 18 - Using the Instrument

Model 14 Oxygen Analyzer5-4TEMP. COMP.MEASURESelectSelectP: 0760.0 mm HgTEMP. COMP.SALINITY COMP.TYPE: ManualTEMP.: 25.02˚C EnterPROGRAMMINGSelectMEAS

Página 19

Model 14 Oxygen Analyzer5-55.1.3.2 ALARMS MenuRelays S1 through S4 may be allocated to the limit, alarm system or timer functions.ALARMSSelectALARM 1

Página 20

Model 14 Oxygen Analyzer5-6➯ SYSTEM ALARM FUNCTION:Relays S3 and S4 may be used as a fault indicater. To control the faults traced by the analyzer, co

Página 21

Model 14 Oxygen Analyzer5-7➯ TIMER FUNCTION:Relays S3 and S4 may be programmed to a timer function.ALARM 4SelectMODE: TimerINTERV: 1440mmIMPUL. Nb.: 5

Página 22 - Programming the Transmitter

Model 14 Oxygen Analyzer5-85.1.3.3 mA OUTPUTS MenuThe analog output signals allow the transmission of the measurements from the analyzer to any extern

Página 23

Model 14 Oxygen AnalyzerTable of ContentsChapter 5 : Programming the Analyzer (Continued) ...

Página 24 - 21.0 ppm

Model 14 Oxygen Analyzer5-9AFFECT - Conc. To set the choice of analog output to the- nA concentration or temperature measurement- ˚C/˚FTYPE - 0/20 C

Página 25 - PROGRAMMING

Model 14 Oxygen Analyzer5-10N° XX Enter number (0 – 32) BAUD - 300 Transmission speed in baud- 600- 1200- 2400- 4800- 9600 PARITY - No Without p

Página 26

Model 14 Oxygen Analyzer5-115.1.4 SERVICE MenuCALIBRATIONAVERAGESOFT ISSUESelectSelectAVERAGE: 9TESTDISPLAYSelectCONC: ppmTEMP: ˚CPRESSURE: mmHgLANGUA

Página 27

5.1.4.1 AVERAGE MenuAn access code may be required.See page 5-14 for CODE menu.mProgram a moving average on the concentration measurement.The measurem

Página 28 - Timer Function

5.1.4.2 DISPLAY MenuDISPLAYCONC: ppmTEMP: ˚CPRESSURE: mmHgLANGUAGE: GBEnterSERVICESelectAVERAGEDISPLAYCODESOFT ISSUEDEFAULT VAL.ADJUST mACONFIGURATION

Página 29 - SPECIAL PROG. OUTPUTS 1/2

5.1.4.3 CODE MenuProtection codes may be programmed for access to the PROGRAMMING, CALIBRATION, and SERVICE menus.This code is deactivated by entering

Página 30 - Special Prog

5.1.4.4 SOFT ISSUE MenuThis menu displays the software version installed in the instrument.LoadingDEFAULT VAL.YesmValues?DefaultEnterSERVICESelectAVER

Página 31

5.1.4.6 mA ADJUST MenuThe analog output signals are factory-adjusted (upper limit: 20mA). However, if the upper limit of one of the outputs, drifts, t

Página 32 - Model 1400

Model 14 Oxygen AnalyzerChapter 6:Calibrating the Instrument6.1. Calibrating the Temperature SensorThe temperature sensor is located in the dissolved

Página 33

Model 14 Oxygen Analyzer6.1.1. Automatic Temperature CompensationThe sensor continuously measures the sample temperature. The concentration values are

Página 34 - LANGUAGE: GB

Model 14 Oxygen AnalyzerChapter1:Introduction1.1 IntroductionThe Model 14 Dissolved Oxygen Analyzer is a single-channel analyzer for the measurement o

Página 35 - SERVICE: 0000

Model 14 Oxygen Analyzer6.1.2. Manual Temperature Compensation Manual temperature compensation should be used only when the sample temperature is cons

Página 36

Model 14 Oxygen Analyzer6.2. Measurement Calibration The slope calibration can be done in air or water. The zero can be set with an electronic zero or

Página 37 - CONFIGURATION CONFIGURATION

Model 14 Oxygen Analyzer6.2.2. Slope Calibration in Air with a Chemical Zero6-5ZERO CALIB.CAL21.2ppm25.0 ˚COKEXECUTIONSelectPROGRAMMINGCAL0ppm25.0 ˚CO

Página 38 - Calibrating the Instrument

Model 14 Oxygen Analyzer6.2.3. Slope Calibration in Process with an Electronic Zero6-6CONC. CALIB.ZERO CALIB.ZERO CALIB. SLOPE CALIB.SLOPE CALIB.Selec

Página 39 - 21.2 21.2

Model 14 Oxygen AnalyzerSLOPE :The flashing “CAL” message indicates that theinstrument is in calibration mode. Wait for the current to stabilize(appro

Página 40

Model 14 Oxygen Analyzer6-8

Página 41

Model 14 Oxygen AnalyzerChapter 7:Start-up and Troubleshooting7.1 Start-up7.1.1 Probe ConnectionSee page 3-3 for the terminal connection. 7.1.2 Main

Página 42

Model 14 Oxygen AnalyzerCauses and solutions:A) The membrane permeability has changed (coating).➯ Calibrate the analyzer and check if the concentratio

Página 43

Model 14 Oxygen Analyzer7.3 Troubleshooting the Electronics7-3WARNING!Never attempt to service before disconnecting the instrument from the main power

Página 44

Model 14 Oxygen Analyzer7-4Interferences Are there any sources of potential interference, chemical, external, temperature, pressure, etc.?Cable shield

Página 45

Model 14 Oxygen Analyzer1.3 Main Characteristics Range: 0 - 299.99% saturation; 0 - 29.99 ppm Calibration in the air Temperature compensation Pr

Página 46 - Start-up and Troubleshooting

Model 14 Oxygen AnalyzerChapter 8: Error MessagesDuring failure, the measurement is replaced by dashes “ - - -.”8-1CONCENTRATIONTOO HIGH- - -STAND-BY1

Página 47

Model 14 Oxygen Analyzer8-2Error messages during calibrationThe current value exceeds the limits. Confirmthere is no short-circuit in the measuring li

Página 48

Model 14 Oxygen AnalyzerAppendix 1: Pressure Conversion TableA-1hPa or mbars mm Hg mm H2O psi inches Hg inches H2O 920.00 690 9503.60 13.34 27.1

Página 49

Model 14 Oxygen AnalyzerAppendix 2: Temperature Conversion TableA-2• Conversion from ˚C into ˚F : ˚F = 1.8 *˚C + 32• Conversion from ˚C into ˚K : ˚K =

Página 50 - Chapter 8: Error Messages

Model 14 Oxygen AnalyzerAppendix 3: Default ValuesCONC. CALIB. PARAMETERSZERO CALIB. DATE :01/01/98Type : Elec. Auto ZERO: 0.00 nASLOPE S: 3.000 nA/pp

Página 51

Model 14 Oxygen AnalyzerA-4AVERAGEAVERAGE : 1 DISPLAYDISPLAYCONC. : ppb/ppmTEMP. : °CPRESSURE : mmHgLANGUAGE : GB CODECODECALIB. : 0000PROG. : 0000SE

Página 52 - Pressure Conversion Table

Model 14 Oxygen AnalyzerA-5Broadley Technologies Corporation19 ThomasIrvine, California 92618 USACorporate Headquarters:Toll-Free: 877.246.7900 Tel:

Página 53 - Temperature Conversion Table

Model 14 Oxygen AnalyzerDisplay Display in concentration units or % saturationDisplay of the direct cell current in nADisplay of the sample temperatur

Página 54 - Appendix 3: Default Values

Model 14 Oxygen Analyzer1.5 Dimensions(Dimensions are in mm and inches)1-4Esc Enter33.5 mm24 mm 0.936"1.319"24 mm 0.945" 28 mm 1.102&qu

Página 55

Model 14 Oxygen AnalyzerChapter2: Description of the Analyzer2.1 Transmitter2.1.1 Presentation of the Transmitter2-1The electronic unit amplifies the

Página 56

Model 14 Oxygen Analyzer2-22.1.2 Transmitter Schematic:1 : Programmable potentiostat 6 : Auxiliary input 2 : Polarization voltage amplifier 7 : Mul