Broadley Technologies 14 Manual de Instruções Página 39

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 38
Model 14 Oxygen Analyzer
6.1.1. Automatic Temperature Compensation
The sensor continuously measures the sample temperature. The concentration values are automatically
calculated using the algorithms programmed in the transmitter.
6-2
TEMP. CALIB.
TEMP. CALIB.
CALCAL
21.2 21.2
˚C ˚C
6.87ppm
OK
Press OK when the
measurement is stable
Can change the value
EXECUTION
T: 0.0 ˚C
ENTER : ok
ESC : cancel
Select
21.2 ˚C
Enter
Enter
1. Programming
Choose automatic temperature compensation. ?
TEMP. COMP.
MEASURE
Select
Select
P: 0760.0 mm Hg
TEMP. COMP
SALINITY COMP.
TYPE: Auto.
Enter
PROGRAMMING
Select
MEASURE
ALARMS
mA OUTPUTS
RS485
Enter
2. Execution
The execution is begun in the CALIBRATION
menu.
Enter the sample temperature value in °C.
Press OK to adjust the temperature displayed
to the actual value of the sample temperature
measured with a precise thermometer.
The gap between the calibration and the
theoretical response curve of the sensor is
given for information.
Vista de página 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 55 56

Comentários a estes Manuais

Sem comentários