Model 14 Oxygen Analyzer
5-7
TIMER FUNCTION:
Relays S3 and S4 may be programmed to a timer function.
ALARM 4
Select
MODE: Timer
INTERV: 1440mm
IMPUL. Nb.: 5
Ton: 005s
Toff: 003s
TmA: 05mn
Relay S4
Closed
Open
Time
pulse number
Ton Toff
TmA
Interval
Example of a timer operating cycle :
MODE -No The Alarm 4 may be a limit (see parameters
-Limit above) or a timer function
-Timer
INTERV XXXXmn Interval between 2 active cycles (in minutes)
IMPUL.Nb: X Number of pulses during an active cycle
Ton XXXs Adjustment of the relay active time (in seconds)
for each pulse
Toff XXXs Adjustment of the relay inactive time (in seconds)
for each pulse
TmA XXmn Hold time for the analog outputs after each cycle
The analog output status depends on the
configuration of the menu mA OUTPUTS/
SPECIAL PROG./TIMER
Timer Function
The measurement cycle lasts 4 seconds.